Alarm Prompt | Cause | LycoVent Action | Clinician Action | Alarm Settings |
---|---|---|---|---|
Encoder Fault | Encoder returns abnormal or unexpected pulse response. | Immediately stop compression, open completely. Resume normal operation at start of breathing cycle. Visible alert on screen and High Priority tone. | Check for normal arm movement Check for obstruction Consult manufacturer. Revert to manual ventilation. | |
Low Battery | Battery level drops below nominal charge. | Continue normal operation. Visible alert on screen and Medium Priority tone. | Connect the LycoVent to mains power supply. | |
Mains Power Disconnected | Ventilator is no longer running from mains power/ switched to battery. | Continue operation on battery. Charge icon no longer visible. Visible alert on screen and Medium Priority tone. | Connect to mains supply when possible. Unit has limited running time on battery and will need to be monitored. | |
Max Motor Load Exceeded | Over Current Fault detected by motor controller. | Immediately stop compression, open completely. Resume normal operation at start of breathing cycle. Visible alert on screen and High Priority tone. | Check for obstructions. If problem persists, consult manufacturer. Revert to manual ventilation. | |
System Fault | Failure of internal system components – motor, sensors, power, switches. | Attempt to continue normal operation Visible alert on screen and High Priority tone. NOTE: Failure of pressure sensor will cease operation of LycoVent | Check for normal operation and observe any concurrent alarms. Consult manufacturer. Revert to manual ventilation. | |
Watch Dog Alarm | Failure of the main controller,its pressure sensor or over-travel of Paddle arms. | Depends on source see Section 4.0 Patient Safety | This is very unusual alarm and implies failure of critical sensor or system. Ventilator will continue to operate if possible to minimise risk to patient till arrival of clinical team. Revert to manual ventilation and contact manufacturer. | |
Zero Speed Fault | Motor shaft motion is not detected by the encoder. | Immediately stop compression, open completely. Resume normal operation at start of breathing cycle. Visible alert on screen and Medium Priority tone. | Check for obstruction Consult manufacturer. Revert to manual ventilation. |