Alarm

Trigger

Clear Condition

Limit Type

Delay

Over Current Fault

Motor current > 3A

Full breathing cycle with motor current below 5A

Upper

500 ms

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Timed Count

3 within 1min

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Nil

Over Current Fault

High load on motor detected. Check for obstruction in mechanism.

Technical