Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Encoder Fault | Encoder reading at known position (switch) deviates by more than 2 degrees | Encoder reading at known position (switch) within 2 degrees | Deviation | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | High | Timed Count | 3 within 5min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Nil | Encoder Fault Arm positioning error. Check for correct movement and ensure no obstructions in mechanism. | Technical |