Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Watchdog Alarm | Watchdog initiated trip – end limit, heartbeat timeout, high pressure safety. | System startup | Detect | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
High | High | N/A | N/A | Yes |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Nil | Watchdog Protection Triggered Watchdog safety activated. Switch to manual ventilation and run system check on ventilator. Do not operate with system check failure. | Technical |