Alarm

Trigger

Clear Condition

Limit Type

Delay

Zero Speed Fault

Motor output with no encoder movement

Full breathing cycle with movement detected

Lower

1 sec

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Consecutive Count

3

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Encoder Fault

Zero Speed Fault

No movement detected when required.

Check for obstruction in mechanism or presence of abnormal noise/smell.

Technical