View Categories

Technical Specifications

< 1 min read

FeaturesLycoVent
Device ClassClass I
Power Source115 or 240 V mains Allowable range 80 to 264 V
Ventilator ModesVolume Controlled Ventilation (VCV),  Assist Control Ventilation(ACV), Suction mode
Altitude CompensationSea level to 2500m factory set. Can be reset remotely by manufacturer.
Ventilator Adjustable Parameters
Tidal Volume Range (mL)200 – 800
PEEP (cm H2O)5-20
Respiratory rate (bpm)8 – 30 (refer Section 8.4)
Inspiratory time (s)0.5 – 2 (refer Section 8.4)
Inspiratory to expiratory time ratio (I:E)1:1 to 1:4 (refer Section 8.4)
FiO2 (%)21 – 100
Alarm Setup Parameters
AlarmsRefer Section 10
Pmax (cm H2O)-Auto ResetDefault 50 and Max
Driving H (cm H2O)Default 35
Plateau H (cm H2O)Default 30
Plateau L (cm H2O)Default 10
TV Range (%)Default 10
Trigger Pressure (ACV mode)Neg 0 to 20 cm H2O
Trigger Flow (ACV mode)0 to 100 mL
FIO2 High21 to 100
FIO2 Low21 to 100
Suction Time0 to 5 mins
Pressure trip Watch Dog  – Auto Reset at reduced TV  if main controller operable70 cm H2O
Ventilation Circuit Connection22 mm outside diameter (OD) ‘male’ standard connection
Battery12.8 V 3.8 Ah Lithium Iron Phosphate type with protection from overcharging, over-discharge, overcurrent, over-temperature and with balancing function
Battery Operating Time4 hours
Dimensions530 x 330 x 200 mm
Weight (kg)16
Motor12 V 48.7 W output rated at 3,000 rpm
Gearbox126:1 ratio 3 stage
Encoder500 ppr
TouchscreenCapacitive Touch Panel
Sound level<70 dbA