Features | LycoVent |
---|---|
Device Class | Class I |
Power Source | 115 or 240 V mains Allowable range 80 to 264 V |
Ventilator Modes | Volume Controlled Ventilation (VCV), Assist Control Ventilation(ACV), Suction mode |
Altitude Compensation | Sea level to 2500m factory set. Can be reset remotely by manufacturer. |
Ventilator Adjustable Parameters | |
Tidal Volume Range (mL) | 200 – 800 |
PEEP (cm H2O) | 5-20 |
Respiratory rate (bpm) | 8 – 30 (refer Section 8.4) |
Inspiratory time (s) | 0.5 – 2 (refer Section 8.4) |
Inspiratory to expiratory time ratio (I:E) | 1:1 to 1:4 (refer Section 8.4) |
FiO2 (%) | 21 – 100 |
Alarm Setup Parameters | |
Alarms | Refer Section 10 |
Pmax (cm H2O)-Auto Reset | Default 50 and Max |
Driving H (cm H2O) | Default 35 |
Plateau H (cm H2O) | Default 30 |
Plateau L (cm H2O) | Default 10 |
TV Range (%) | Default 10 |
Trigger Pressure (ACV mode) | Neg 0 to 20 cm H2O |
Trigger Flow (ACV mode) | 0 to 100 mL |
FIO2 High | 21 to 100 |
FIO2 Low | 21 to 100 |
Suction Time | 0 to 5 mins |
Pressure trip Watch Dog – Auto Reset at reduced TV if main controller operable | 70 cm H2O |
Ventilation Circuit Connection | 22 mm outside diameter (OD) ‘male’ standard connection |
Battery | 12.8 V 3.8 Ah Lithium Iron Phosphate type with protection from overcharging, over-discharge, overcurrent, over-temperature and with balancing function |
Battery Operating Time | 4 hours |
Dimensions | 530 x 330 x 200 mm |
Weight (kg) | 16 |
Motor | 12 V 48.7 W output rated at 3,000 rpm |
Gearbox | 126:1 ratio 3 stage |
Encoder | 500 ppr |
Touchscreen | Capacitive Touch Panel |
Sound level | <70 dbA |
Technical Specifications