Measurement Accuracy

The accuracy of parameters displayed on-screen was tested using calibrated instruments and the % error for each parameter was as follows:

Display Pressures (Claimed accuracy 5.5% or better)

  • 5 cm H2O ± 5.5%
  • 37 cm H2O ± 2.0%
  • 50 cm H2O ± 1.3% (by extrapolation)

Display Volumes (Claimed accuracy 4% or better)

VexpMV
205 mL  ± 3.4%2,795 mL ± 3.6%
484 mL  ± 1.4%6,305 mL ± 3.7%
721 mL  ± 0.7%9,336 mL ± 1.9%
800 mL  ± 0.6% (by extrapolation)

Display Oxygen (Claimed accuracy 3% or better)

  • 30% ± 2 %
  • 60% ± 1%
  • 90% ± 1%

FIO2 System Response Time

The response times (See Overview of the LycoVent: Oxygen Level Control) are based on the time to achieve 90% of a target FiO2 of 60% from a starting FiO2 of 22% (air) through the addition of medical grade O2 (99%).