Alarm

Trigger

Clear Condition

Limit Type

Delay

Inspiratory Delivery High

Measured Inspired volume > 1.1 x TV SP

Measured Inspired volume < 1.1 x TV SP

Upper

First breathing cycle

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

Medium

Timed Count

3 within 1min

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Flow Senor Fault

Inspiratory Delivery High

Measured flow delivered exceeds set tidal volume. Confirm tidal volume and check for correct flow to patient.

Technical