Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Inspiratory Delivery High | Measured Inspired volume > 1.1 x TV SP | Measured Inspired volume < 1.1 x TV SP | Upper | First breathing cycle |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | Medium | Timed Count | 3 within 1min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Flow Senor Fault | Inspiratory Delivery High Measured flow delivered exceeds set tidal volume. Confirm tidal volume and check for correct flow to patient. | Technical |