Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Expiratory Flow Low | Measured Expired volume < 0.9 x Measured Inspired volume | Measured Expired volume > 0.9 x Measured Inspired volume | Lower | First breathing cycle |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | High | Timed Count | 3 within 1min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Flow Senor Fault | Expiratory Flow Low Measured expiratory flow less than delivery. Inspect for leaks and check for correct flow from patient. | Technical |