Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Low Tidal Pressure | Peak pressure – PEEP < TPres** | Peak pressure – PEEP > TPres** | Lower | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | Medium | N/A | N/A | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Pressure Sensor Fault | Low Tidal Pressure Low pressure detected for expected tidal volume. Check bag is seated correctly in ventilator cradle. | Physiological |