Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Pressure Sensor Fault | Pressure not detected when expected or no change in pressure. | Pressure detected within range. | Detect | 5 sec |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
High | High | N/A | N/A | Yes |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Nil | Pressure Sensor Fault Error in pressure measurement. Potential sensor failure. Refer to technician if problem persists. | Technical |