Alarm

Trigger

Clear Condition

Limit Type

Delay

High PEEP

PEEP > 1.1 x PEEP setpoint

PEEP < PEEP setpoint

Upper

First breathing cycle

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Low

High

Timed Count

5 min

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

0 – 20 cm H₂O

Pressure Sensor Fault

High PEEP

PEEP too high. Adjust manual PEEP valve accordingly and ensure complete expiration before next breath.

Technical