Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Tidal Volume Not Reached | Encoder position at inspiratory pause < 0.9xTidalPos*** | Encoder position at inspiratory pause > 0.9xTidalPos*** | Lower | First breathing cycle |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Low | Medium | Consecutive Count | 5 | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
Yes | 0 – 50 % | Encoder Fault | Tidal Volume Not Reached Desired tidal volume not delivered. Observe system for adequate ventilation and adjust as required. | Technical |