Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
FiO2 High | Measured FiO2 exceeds FiO2 setpoint by 10% | Measured FiO2 within 10% of FiO2 setpoint | Deviation | 3 min on change |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | High | Timed | 5 min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
Yes | 21 - 100% | FiO2 Sensor Fault | FIO2 High FIO2 is too high. Check oxygen supply and adjust as required. | Physiological |