Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Ventilator Stopped | Ventilator stopped with positive PEEP or within detected breathing cycle. | Ventilator run | Detect | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
High | High | N/A | N/A | Yes |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Nil | Ventilator Stopped Ventilation was ceased while patient connected. Ensure ventilation is no longer required or restart ventilator. | Physiological |