Alarm

Trigger

Clear Condition

Limit Type

Delay

Watchdog Alarm

Watchdog initiated trip – end limit, heartbeat timeout, high pressure safety.

System startup

Detect

Nil

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

High

High

N/A

N/A

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Nil

Watchdog Protection Triggered

Watchdog safety activated. Switch to manual ventilation and run system check on ventilator.

Do not operate with system check failure.

Technical