Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
System Fault | Internal system faults – voltage faults, watchdog fault. Instrument failure. | No internal system faults | Detect | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
High | High | N/A | N/A | Yes |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Nil | System Fault Manually ventilate patient. Refer ventilator to technician. | Technical |