Alarm

Trigger

Clear Condition

Limit Type

Delay

Low Battery

Battery voltage < 12.4V

Battery voltage > 12.4V

Lower

1 min

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Timed or Measured

> 30 min or battery voltage < 12V

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Nil

Low Battery

Low charge on battery. Plug into mains to recharge.

Technical