Alarm

Trigger

Clear Condition

Limit Type

Delay

FiO2 Sensor Fault

Voltage over-range or under-range

Measurement within range

Detect

1 min on change

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Timed

5 min

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

N/A

Nil

FIO2 Sensor Fault

Error in FIO2 measurement. Potential sensor failure. Refer to technician if problem persists.

Technical