Alarm

Trigger

Clear Condition

Limit Type

Delay

FiO2 High

Measured FiO2 exceeds FiO2 setpoint by 10%

Measured FiO2 within 10% of FiO2 setpoint

Deviation

3 min on change

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Timed

5 min

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

21 - 100%

FiO2 Sensor Fault

FIO2 High

FIO2 is too high. Check oxygen supply and adjust as required.

Physiological