Alarm

Trigger

Clear Condition

Limit Type

Delay

Tidal Volume Not Reached

Encoder position at inspiratory pause < 0.9xTidalPos***

Encoder position at inspiratory pause > 0.9xTidalPos***

Lower

First breathing cycle

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Low

Medium

Consecutive Count

5

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

0 – 50 %

Encoder Fault

Tidal Volume Not Reached

Desired tidal volume not delivered. Observe system for adequate ventilation and adjust as required.

Technical