Alarm

Trigger

Clear Condition

Limit Type

Delay

Expiratory Flow Low

Measured Expired volume < 0.9 x Measured Inspired volume

Measured Expired volume > 0.9 x Measured Inspired volume

Lower

First breathing cycle

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Medium

High

Timed Count

3 within 1min

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Flow Senor Fault

Expiratory Flow Low

Measured expiratory flow less than delivery. Inspect for leaks and check for correct flow from patient.

Technical