Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Minute Volume High | Minute Volume > spTV x spRR x 1.15 | Minute Volume < spTV x spRR x 1.15 | Upper | 1min |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Low | High | Timed | 5 min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
Yes | 0 – 25 L | Flow Senor Fault | Minute Volume High Excessive ventilation detected. Observe patient and adjust parameters as appropriate. | Technical |