Alarm

Trigger

Clear Condition

Limit Type

Delay

Pressure Sensor Fault

Pressure not detected when expected or no change in pressure.

Pressure detected within range.

Detect

5 sec

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

High

High

N/A

N/A

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

No

N/A

Nil

Pressure Sensor Fault

Error in pressure measurement. Potential sensor failure. Refer to technician if problem persists.

Technical