Alarm

Trigger

Clear Condition

Limit Type

Delay

High Plateau Pressure

Plateau pressure > PlatH*

Full breathing cycle with Plateau pressure below PlatH*

Upper

Nil

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

High

High

N/A

N/A

Yes

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

PlatL ➡ 50 cm H₂O

Pressure Sensor Fault

High Plateau Pressure

Check patient and confirm lung pressures are within acceptable limit.

Physiological