Alarm

Trigger

Clear Condition

Limit Type

Delay

Low Pressure Disconnect

Plateau pressure < PlatL*

Full breathing cycle with plateau pressure above PlatL*

Lower

Nil

Priority

Max Escalation

Escalation Type

Escalation Condition

Latching

Low

Medium

Timed

1 min

No

User Adjustable

Adjustment Range

Alarm Masking

Tooltip Message

Physiological/Tech

Yes

0 – PlatH cm H₂O

Pressure Sensor Fault

Low Pressure Disconnect

Potential disconnection detected in breathing circuit. Check breathing circuit for potential leaks.

Physiological