Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Leak Detected | PEEP < 0.9 * PEEP setpoint | PEEP > PEEP setpoint | Lower | First breathing cycle |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
High | High | N/A | N/A | Yes |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
Yes | 0 – 20 cm H₂O | Pressure Sensor Fault | Leak Detected Leak or disconnect detected in system. Check breathing circuit for potential leaks. | Technical |