Alarm | Trigger | Clear Condition | Limit Type | Delay |
---|---|---|---|---|
Driving Pressure High | (Peak pressure – PEEP) > (Plateau pressure – PEEP) x 2 | Full breathing cycle within driving pressure upper limit | Upper | Nil |
Priority | Max Escalation | Escalation Type | Escalation Condition | Latching |
Medium | High | Timed | 1 min | No |
User Adjustable | Adjustment Range | Alarm Masking | Tooltip Message | Physiological/Tech |
No | N/A | Pressure Sensor Fault | Driving Pressure High Potential restriction between bellows and patient. Check for obstruction or kinks in breathing circuit. | Physiological |