Alarm Prompt | Cause | LycoVent Action | Clinician Action | Alarm Settings |
---|---|---|---|---|
Abnormal Plateau Pressure | Measured plateau pressure is higher than the measured peak pressure. | Continue normal operation. Visible alert on screen and Medium Priority tone. | Check for patient cough Revert to manual ventilation if required. | |
Driving Pressure Hi | Peak Pressure - PEEP is larger than (Plateau Press – PEEP) x 2 | Continue normal operation. Visible alert on screen and High Priority tone tone. | Check Driving H value set on alarm setup is the required value. Default value is 35 cm H2O. Confirm PEEP valve value is the same as PEEP setting on LycoVent. Check the patient circuit and BMV for any obstructions that could be adding resistance.
| |
Hi PEEP | Measured PEEP value is more than 1.1 x PEEP set value | Continue normal operation. Visible alert on screen and Medium Priority tone | Check the PEEP value set on the LycoVent matches the value set on the PEEP valve. Check for pressure stacking and adjust ventilation parameters accordingly. Revert to manual ventilation if required. | |
Leak Detected | Measured PEEP value is less than 0.9 x PEEP set value. | Continue normal operation. Visible alert on screen and High Priority tone.
| Check the PEEP value set on the LycoVent matches the value set on the PEEP valve. Inspect PEEP valve for any signs of wear and ensure it is connected to the patient circuit correctly. Check the patient circuit for leaks. Revert to manual ventilation if required. | |
Low Pressure Disconnect | Plateau pressure is lower than the Plat L set on the ventilator. | Continue normal operation. Visible alert on screen and Low Priority tone. | Check Plateau L value set on alarm setup is the required value. Correct if required. Check patient circuit and BMV for any leaks or disconnects. Confirm ventilation parameters. Correct if required. | |
Plateau Pressure High | Measured plateau pressure is higher than the set Plateau H value on the LycoVent. | Immediately stop compression, open completely. Resume normal operation at start of breathing cycle. Visible alert on screen and High Priority tone | Check the Plateau H value set on the LycoVent is the required value. Default is set to 30 cm H2O. Confirm ventilation settings. Correct if required Revert to manual ventilation if required. | |
PMAX Exceeded | The inspiratory pressure is higher than maximum allowed pressure set on ventilator- 50cmH2O
| Immediately stop compression, open completely. Resume normal operation at start of breathing cycle. Visible alert on screen and High Priority tone. | Attend to patient and revert to manual ventilation if required. Confirm the PMAX value set on the alarm setup is the required value. Correct if required. Confirm ventilation settings. Correct if required. Inspect patient breathing circuit and BMV for obstructions / faults. Replace component if required. | |
Pressure Tx Fault | Pressure not detected or no change | LycoVent will continue operation. Visible Alert on screen High Priority tone | Check sensor tubing correctly connected. If problem persists, revert to manual ventilation until alternative ventilator connected. Contact manufacturer | |
Tidal Press Lo | Tidal pressure is measured below 10 cm H2O. | Continue normal operation. Visible alert on screen and Medium Priority tone. | Check for correct configuration and/or leaks |